S9130 Sleep Therapy Apnea BIPAP Machine Non-invasive Ventilator
Main technical parameters of S9130 non-invasive ventilator
1. Performance characteristics
1. Use a 4.3-inch color LCD screen to display breathing parameters, alarm information and waveform diagrams;
2. The expanded 15.6-inch display can be rotated by 90 degrees, which is convenient for medical staff to observe in multiple directions, and can directly display the flow waveform and pressure waveform;
3. It is suitable for non-invasive respiratory ventilation patients with respiratory insufficiency, and can be used in home and medical environments.
4. High-performance imported fan, compact body, and the equipment comes with a humidifier to increase air humidity;
5. Optional trolley to facilitate the transfer between hospital beds;
6. Optional humidifier to heat and humidify the inhaled gas, improve comfort, and effectively prevent complications caused by obstruction caused by dry airway;
7. Possess the National Class III Medical Device Registration Certificate;
8. Standard lithium battery with a capacity of 28500mAh, which can work independently and continuously for ≥8 hours, and can work continuously for ≥2 hours with an external expansion screen;
9. Low pressure oxygen can be connected externally, and the oxygen concentration range is 30-50%;
10. Noise ≤45dB(A) under 10cmH2O pressure level.
2. Working conditions
1. Power supply: DC 24 V (-15 to +25 %), 2.5 A
2. Power adapter: AC input 100~240 V, frequency 50/60 HZ DC output 24 V, 2.5 A
3. Ambient temperature range +5 ~ +35 ºC
4. The relative humidity range is 15 ~ 95%
5. Atmospheric pressure range 830 ~ 1060 hPa
3. Breathing mode and function
1. Spontaneous breathing and time control automatic switching mode (S/T mode)
2. Time control mode (T mode)
3. Spontaneous breathing mode (S mode)
4. Continuous positive pressure ventilation mode (CPAP mode)
5. Pressure control mode (PC)
6. Pressure automatic titration function (AUTO)
7. Average capacity guarantee pressure support function (AVAPS)
Ventilation mode: S,T,S/T,PC,AVAPS,CPAP,AUTO
Fourth, the main technical indicators
Tuning parameters
1. AVPAS on/off
2. Tidal volume 200 ~ 1500 mL step: 10 mL
3. Breathing rate 3 ~ 40 bpm step: 1 bpm
4. Inhalation time 0.0 ~ 3.0 S step: 0.1 S
5. Inspiratory pressure (IPAP) 4 ~ 30 cmH2O step: 0.5 cmH2O
6. Maximum suction pressure (IPAPMAX) 4 ~ 30 cmH2O step: 0.5 cmH2O
7. Minimum suction pressure (IPAPMIN) 4 ~ 30 cmH2O step: 0.5 cmH2O
8. Expiratory pressure (EPAP) 4 ~ 25 cmH2O step: 0.5 cmH2O
9. Continuous positive airway pressure (CPAP) 4 ~ 20 cmH2O step: 0.5 cmH2O
10. Pressure delay rise time 0 ~ 60 min Step: 1 min
11. The pressure rise time is automatically adjusted, 0.1 S ~ 0.6 S step: 0.1 S
12. Inhalation sensitivity automatic adjustment, 1, 2, 3 step: 1
13. Exhalation sensitivity automatic adjustment, 1, 2, 3 step: 1
14. Comfort level off, 1, 2, 3 step: 1
15. Humidifier 0 ~ 5 step: 1
16. Automatic function on/off
17. Pressure limit 20 ~ 35 cmH2O step: 0.5 cmH2O
Monitoring parameters
1. Frequency (Freq) 0 ~ 100 bpm Resolution: 1 bpm
2. Tidal volume (VT) 0 ~ 2000 mL Resolution: 1 mL
3. Minute ventilation (MV) 0 ~ 99 L/min ≥10 L/min, 0.1 L/min; <10 L/min, 0.01 L/min
4. Inspiratory time (Time) 0.0 ~ 3.0 S Resolution: 0.1 S
5. Inspiratory pressure (IPAP) 0 ~ 40 cmH2O Resolution: 0.5 cmH2O
6. Expiratory pressure (EPAP) 0 ~ 40 cmH2O Resolution: 0.5 cmH2O
7. Positive airway pressure (CPAP) 4 ~ 20 cmH2O Resolution: 0.5 cmH2O
Five, graphic display
1. Airway pressure-time waveform
2. Flow-time waveform
6. Alarm and protection
1. Asphyxia alarm [Asphyxia lasts for 8 s~12 s, and there is no exhalation]
2. Pipe falling off alarm [When the breathing tube is disconnected, and it lasts for 16s]
3. Alarm system silence (silence) [Alarm silence time: ≤120 s]
4. Low ventilation alarm [range: off, 0 L/min~15 L/min]
5. High breathing rate alarm Range: Off, 1 bpm ~ 99 bpm
6. Low breathing rate alarm Range: Off, 1 bpm ~ 60 bpm
7. Airway high pressure alarm Range: 20 cmH2O ~ 35 cmH2O
8. Power failure alarm
9. Low tidal volume alarm [When AVAPS is turned on]