• Turbine Ventilator Driven High Frequency Electric Control ICU Infant Neonatal Ventilator Respirator
  • Turbine Ventilator Driven High Frequency Electric Control ICU Infant Neonatal Ventilator Respirator
  • Turbine Ventilator Driven High Frequency Electric Control ICU Infant Neonatal Ventilator Respirator
  • Turbine Ventilator Driven High Frequency Electric Control ICU Infant Neonatal Ventilator Respirator
  • Turbine Ventilator Driven High Frequency Electric Control ICU Infant Neonatal Ventilator Respirator

Turbine Ventilator Driven High Frequency Electric Control ICU Infant Neonatal Ventilator Respirator

Type: Ventilator
Kind: Ventilator Machine
Certification: CE
LCD Display: With LCD Display
Certificate: CE
Transport Package: Wooden Case
Samples:
US$ 8300/Piece 1 Piece(Min.Order)
| Request Sample
Customization:
Diamond Member Since 2018

Suppliers with verified business licenses

Manufacturer/Factory & Trading Company

Basic Info.

Model NO.
S1100C
Specification
560*560*605mm /35KG 670*700*1160mm /80KG
Origin
China

Product Description

Features

1,15" TFT touch screen displays the ventilation parameters, alarm information, and oscillograms, making every operation more easily. 2, Multiple ventilation modes, suitable for ICU, emergency department and operation room, etc: (PRVC, APRV, DUOLEVEL, V-SIMV, P-SIMV, V-A/C, V-A/C, IPPV, PCV, PSV, SPONT/CPAP, SIGH, MANUAL). 3,5 oscillograms for your choice, 3 of them can be displayed on the screen at the same time.
 
4, Humidifiers can heat and wet breathing gas, making it comfortable for patients to breathe. 5, Rapid oxygen supply automatically offers a high flow rate of oxygen within two minutes. 6, Nebulization makes medicine into small liquids and particles,easier and quicker for the patient to breathe in. 7, High-temperature resistance breathing circuit is reusable and anti-pollution.

Application

The ventilator makes a good performance in the operation room, ICU department, and emergency treatment. It is used to assist or replace spontaneous breathing for adult pediatric and neonatal patients. 25 years of experience in Market-oriented ventilators make us professional and reliable, satisfying all your needs in ventilation. Due to the flexible configuration, good quality, and competitive price, S1100 has soon become the superstar of the market.
 

Safety

1,14 types of sound and visual alarm information, easier for users to do some error checking and troubleshooting. 2, Built-in oxygen concentration sensor, ensures stable precision of oxygen concentration. 3, Easy to move with four casters, easy to stop with two brakes. 4, Separate design of electronic circuit and gas flow rate keep safe running of the ventilator. 5, Compact long-life internal battery can provide emergency power, avoiding the risk to the patient. 6, Self-check before the operation, to eliminate system mistakes.
Turbine Ventilator Driven High Frequency Electric Control ICU Infant Neonatal Ventilator RespiratorTurbine Ventilator Driven High Frequency Electric Control ICU Infant Neonatal Ventilator Respirator
Specification
Ventilation mode
PRVC, APRV, DUOLEVEL, V-SIMV, P-SIMV, P-A/C, V-A/C, IPPV, PCV, PSV, SPONT/CPAP, SIGH, MANUAL
Ventilator parameter range
Tidal volume (Vt)
0, 20 ~ 2000 mL
Frequency (Freq)
1 min ~ 100 min
Oxygen concentration
21 % ~ 100 %
I: E
4: 1 ~ 1: 8
PEEP
0 cmH2O ~ 40 cmH2O
Pressure triggering sensitivity (Ptr)
-20 cmH2O ~ 0 cmH2O (Based on PEEP)
Flow trigger sensitivity (Ftr)
0.5 L/min ~ 30 L/min
Pressure control (PC)
5 cmH2O ~ 80 cmH2O
Pressure support (PS)
0 cmH2O ~ 80 cmH2O
SIGH
0 (off) 1/100 ~ 5/100
Apnea ventilation
OFF, 5 s ~ 60 s
Pressure limit
20 cmH2O ~ 100 cmH2O
Oscillogram
P-T(Pressure-Time) 
F-T(Flow-Time) 
V-T(Volume-Time) 
P-V Loop(Pressure-Volume Loop) 
F-V Loop(Flow-Volume Loop) 
P-F Loop(Pressure-Flow Loop)
Alarm and protection
The AC power failure alarm
Power failure or no connection
Internal battery backup low voltage alarm
≤ 11.3 ± 0.3 V
No tidal volume
No tidal volume within 6 s
High oxygen concentration alarm
19 % ~ 100 %
Low oxygen concentration alarm
18 % ~ 99 %
High airway pressure alarm
20 cmH2O ~ 100 cmH2O
Low airway pressure alarm
0 cmH2O ~ 20 cmH2O
High-minute volume alarm
5 L/min ~ 99 L/min
Low-minute volume alarm
1 L/min ~ 30 L/min
Continuous pressure alarm
(PEEP + 1.5 cmH2O) over 16s
Suffocation warning
5 s ~ 60 s 
The maximum limited pressure
≤12.5 kPa
Fan error
Show on screen
Oxygen deficit
Show on screen
Working condition
Gas source
O2, Air
Pressure
280 kPa ~ 600 kPa
Voltage
220 V ± 22 V
Power frequency
50 Hz ± 1 Hz
Input power
900 VA(With air compressor) 
250 VA(Without air compressor)
Monitoring parameter
Frequency(Freq)
0 /min ~ 100 /min
Tidal volume (Vt)
0 mL ~ 2500 mL
MV
0 L/min ~ 99 L/min
Airway pressure
0 cmH2O ~ 100 cmH2O
Dynamic lung compliance monitoring
1 mL/cmH2O ~ 1000 mL/cmH2O
Oxygen concentration
15 % ~ 100 %
Wooden case packing size
Wooden case packing size: L 560 * W 560 * H 615 mm
G.W. :32 KG N.W. :19 KG
 
 
 
 

 

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now

You Might Also Like

Diamond Member Since 2018

Suppliers with verified business licenses

Manufacturer/Factory & Trading Company
Number of Employees
215
Year of Establishment
2008-01-18