Customization: | Available |
---|---|
Type: | Anesthesia Ventilator |
Used for: | Anesthesia Machine |
Still deciding? Get samples of US$ 2300/Piece
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Item No: | DM6B |
Ventilation mode | |
IPPV, A/C, PCV, SIMV, SIGH, MANUAL | |
Ventilator parameter range | |
Flowmeter | O2 (0.1 ~ 10 L/min) |
Rapid oxygen supply | 25 L/min ~ 75 L/min |
Tidal volume(Vt) | 20 mL ~ 1500 Ml |
Frequence(Freq) | 1 /min ~ 100 /min |
I:E | 2:1 ~ 1:6 |
Pressure triggering sensitivity(PTr) | -20 cmH2O~0 cmH2O (Based on PEEP) |
Flow trigger sensitivity (FTr) | 0.5 L/min ~ 30 L/min |
Pressure control (PC) | 5 cmH2O ~ 60 cmH2O |
SIGH | 0 (off) 1/100 ~ 5/100 |
Apnea ventilation | OFF, 5 s ~ 60 s |
Pressure limit | 20 cmH2O ~ 100 cmH2O |
Monitoring parameter | |
Frequence(Freq) | 0 /min ~ 100 /min |
Tidal volume (Vt) | 0 mL ~ 2000 mL |
MV | 0 L/min ~ 100 L/min |
Oscillogram | |
P-T (pressure - time) | |
F-T (flow - time) | |
Alarm and protection | |
The AC power failure alarm | Power failure or no connection |
Low voltage alarm for back up battery | < 11.3 ± 0.3 V |
No tidal volume | ≤ 5 mL within 6 s |
High oxygen concentration alarm | 19% ~ 100% |
Low oxygen concentration alarm | 18% ~ 99% |
High airway pressure alarm | 20cmH2O ~ 100cmH2O |
Low airway pressure alarm | 0cmH2O ~ 20cmH2O |
High minute volume alarm | Adult (5 L/min ~ 20 L/min) |
Low minute volume alarm | Paediatric (1L/min-15L/min) (0-10L/min) |
Continuous pressure alarm | (PEEP+1.5kPa) over 16s |
Suffocation warning | 5s-60s no spontaneous ventilation |
The maximum limited pressure | <12.5 kPa |
Fan error | Show on screen |
Oxygen deficit | Show on screen |
Working condition | |
Gas source | O2 |
Pressure | 280 kPa ~ 600 kPa |
Voltage | 100 ~ 240 V |
Power frequency | 50/60 Hz |