Customization: | Available |
---|---|
Type: | Anaesthesia Machine |
Kind: | Gynaecology |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Ventilation modes: | (V-CMV)IPPV; (P-CMV) PCV; PSV; (P-0SIMV);(V-SIMV); SIGH, PRVC, MANUAL |
Ventilator parameter ranges | |
Graphical display: | |
P-T(pressure - time) | ETCO2-T |
F-T(flow - time) | V-T |
P-V loop (pressure - volume loop) | F-P,F-V |
Alarm and protection | |
The AC power failure alarm | Power failure or no connection |
Internal battery backup low voltage alarm | <<>11.3±0.3V |
No tidal volume | ≤5Ml within 6s |
High oxygen concentration alarm Low oxygen concentration alarm |
19%-100% 18%-99% |
High Airway pressure alarm Low Airway pressure alarm High Minute Volume alarm Low Minute Volume alarm Continuous Pressure alarm |
20cmH2O-100cmH2O 0cmH2O-20cmH2O Adult(5L/min-20L/min) Paed(1L/min-15L/min 0-10L/min) (PEEP+1.5kPa) over 16s |
Suffocation warning | 5s-60s no spontaneous ventilation |
The maximum limited pressure | <<>12.5 kPa |
Fan error Oxygen deficit |
Show on screen Show on screen |
Working conditions | |
Gas source | O2,N2O,AIR |
Pressure | 280kPa-600kPa |
Voltage | 100-240V |
Power frequency | 50/60Hz |
Input power | 85VA |
Vaporizer | 2 standard |
Features