SPECIFICATIONS FOR NLF-200B VENTILATOR
|
Ventilation modes:
|
IPPV, A/C, PCV, SIMV, SPONT/CPAP, PSV
|
||
|
Tidal volume(Vt)
|
0 ~ 2000mL
|
|
Frequency (Freq)
|
1bpm ~ 100 bpm
|
||
Oxygen concentration
|
21%-100%
|
||
I/E
|
4:1~1:8
|
||
PEEP
|
0cmH2O ~ 40 cmH2O
|
||
Pressure Limit
|
20 cmH2O ~ 100 cmH2O
|
||
|
Frequency (Freq)
|
0 /min ~ 100 /min
|
|
Tidal volume(Vt)
|
0 mL ~ 2500 mL
|
||
MV
|
0 L/min ~ 99 L/min
|
||
Airway pressure
|
1mL/cmH2O ~ 1000mL/cmH2O
|
||
Dynamic lung compliance testing
|
1mL/cmH2O ~ 1000mL/cmH2O
|
||
ETCO2 Concentration
|
0mmHg ~ 152mmHg (0%-20%)
|
||
Oxygen concentration
|
15 % ~ 100 %
|
||
ETCO2 Concentration
|
0mmHg ~ 152mmHg (0%-20%)
|
||
Oxygen concentration
|
15 % ~ 100 %
|
||
|
P-T(pressure - time)
|
||
F-T(flow - time)
|
|||
P-V loop (pressure - volume loop)
|
|||
V-T(volume - time )
|
|||
ETCO2-T (End-tidal CO2 - time)
|
|||
|
The AC power failure alarm
|
Power failure or no connection
|
|
Internal battery backup low voltage alarm
|
≤11.3±0.3V
|
||
No tidal volume
|
No tidal volume within 6s
|
||
High Minute Volume alarm
|
5L/min-99L/min
|
||
Low Minute Volume alarm
|
1L/min-30L/min
|
||
High Airway pressure alarm
|
20cmH2O-100cmH2O
|
||
Low Airway pressure alarm
|
0cmH2O-20cmH2O
|
||
High oxygen concentration alarm
|
19%-100%
|
||
Low oxygen concentration alarm
|
18%-99%
|
||
Continuous pressure alarm
|
(PEEP+1.5cmH2O) over 16s
|
||
Suffocation warning
|
5-60s
|
||
Fan error
|
Show on screen
|
||
Oxygen deficit
|
Show on screen
|
||
The maximum limited pressure
|
<12.5 kPa
|
||
|
Gas source
|
O2, AIR
|
|
Pressure
|
280kPa-600kPa
|
||
Voltage
|
220V±22V
|
||
Power frequency
|
50Hz ± 1Hz
|
||
|
900VA (with air compressor)
|
||
250VA (without air compressor)
|
|||
Parking size
|
Main components: L 560 * W 560 * H 605 mm, G.W. : 35 KG, N.W. : 17 KG
|
||